Turbidity Measurement Device
(0 reviews)
Get Price
(89) Views
(0) Contacts

Turbidity measurement device WGZ-20 Series dispersive optical turbidity gauges are used to measure the degree of light scattering from insoluble particulate matter suspended in water or transparent liquids, and to characterize the problem of suspended grain content in these substances.


Contact Company

Hangzhou Lohand Biological Technology Co.,Ltd
Zhejiang, China
(976) Views
(16) Contacts

Turbidity Measurement Device Overhaul:
Failure phenomenon
Possible causes
Repair method
1, after starting up, display
A、power line and socket connection is bad or loose
A、fastening socket or for a power line
2, measuring reaction
A light damage
B, internal connector lose
C, electrical system fault
A, replacement
B, fastening
C, maintenance
3, measurement instability or drift
A, a solution with bubbles or particles in does not stop moving
B, circuitry within the apparatus of damp
C, the outer surface of the bottled water sample
D, external interference
E, the supply voltage instability of
A re-sampling or extended reading time
B, prolonging start preheat time preheating   dehumidification
C, dry specimen bottle
D, eliminating the interference source
E, exclusion of unstable factors
4. zero, less  than zero
A adjustable zero without the use of zero turbidity water
B, zero potentiometer damage
C, zero range migration
D, light path offset
A should adopt zero turbidity water
B, replacement potentiometer
C, W2 potentiometer adjustment board
D, adjustment
5, not to the correction value
A, standard solution standard value is not accurate
B, correction potentiometer is damaged
C, range migration correction
D, light path offset
A, accurate preparation of standard solution
B, replacement
C, W1 potentiometer adjustment board
D, adjustment
6. short stroke
A, due to the transport screen vibration caused by   poor contact
B, LCD LCD screen damage
C, circuit board fault
A, check
B, replacement
C, maintenance

Product name turbidity measurement device
Model WGZ-20
Minimum Readout ( NTU) 0.1
Measuring Range(NTU) 0--200
Basic Error(F.S) ±2.5%
Repeatability ≤1.5%
Zero Draft(F.S) ±0.2%
Power Supply AC220V/50Hz